Acquire:segmented:count - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

5
Commands by Subsystem

:ACQuire:SEGMented:COUNt

Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 250 in NR1 format
This command is available when the segmented memory option (Option SGM) is enabled.
N O T E
The :ACQuire:SEGMented:COUNt command sets the number of memory
segments to acquire.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments in the
current acquisition is returned by the :WAVeform:SEGMented:COUNt?
query.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 250 in NR1 format
See Also
Example Code
172
(see
page
658)
":ACQuire:MODE"
on page 169
":DIGitize"
on page 132
":SINGle"
on page 158
":RUN"
on page 156
":WAVeform:SEGMented:COUNt"
":ACQuire:SEGMented:ANALyze"
"Introduction to :ACQuire Commands"
"Example Code"
on page 173
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
on page 519
on page 171