:OPEE (Operation Status Enable Register)

Command Syntax
:OPEE <mask>
<mask> ::= 16-bit integer
The :OPEE command sets a mask in the Operation Status Enable register.
Set any of the following bits to "1" to enable bit 7 in the Status Byte
Register and potentially cause an SRQ (Service Request interrupt to be
generated.
Table 40 Operation Status Enable Register (OPEE)
Bit
15-13
12
11
10
9
8-6
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
Name
Description
---
---
HWE
Hardware Event
OVLR
Overload
---
---
MTE
Mask Test Event
---
---
Commands by Subsystem
When Set (1 = High = True), Enables:
(Not used.)
Event when hardware event occurs.
Event when 50Ω input overload occurs.
(Not used.)
Event when mask test event occurs.
(Not used.)