Agilent Technologies InfiniiVision 5000 Series Programmer's Manual page 625

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

9
Status Reporting
The bits in the status byte act as summary bits for the data structures
residing behind them. In the case of queues, the summary bit is set if the
queue is not empty. For registers, the summary bit is set if any enabled
bit in the event register is set. The events are enabled with the
corresponding event enable register. Events captured by an event register
remain set until the register is read or cleared. Registers are read with
their associated commands. The *CLS command clears all event registers
and all queues except the output queue. If you send *CLS immediately
after a program message terminator, the output queue is also cleared.
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
625