:TRIGger:CAN:TRIGger

Command Syntax
:TRIGger:CAN:TRIGger <condition>
<condition> ::= {SOF | DATA | ERRor | IDData | IDEither | IDRemote |
The :TRIGger:CAN:TRIGger command sets the CAN trigger on condition:
• SOF - will trigger on the Start of Frame (SOF) bit of a Data frame,
• DATA - will trigger on CAN Data frames matching the specified Id,
• ERRor - will trigger on CAN Error frame.
• IDData - will trigger on CAN frames matching the specified Id of a
• IDEither - will trigger on the specified Id, regardless if it is a Remote
• IDRemote - will trigger on CAN frames matching the specified Id of a
• ALLerrors - will trigger on CAN active error frames and unknown bus
• OVERload - will trigger on CAN overload frames.
• ACKerror - will trigger on a data or remote frame acknowledge bit that
The table below shows the programming parameter and the corresponding
front- panel softkey selection:
Remote <condition> parameter
SOF
DATA
ERRor
IDData
IDEither
IDRemote
ALLerrors
OVERload
ACKerror
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
ALLerrors | OVERload | ACKerror}
Remote Transfer Request (RTR) frame, or an Overload frame.
Data, and the DLC (Data length code).
Data frame.
frame or a Data frame.
Remote frame.
conditions.
is recessive.
Commands by Subsystem
Front-panel Trigger on: softkey selection
(softkey text - softkey popup text)
SOF - Start of Frame
Id & Data - Data Frame Id and Data
Error - Error frame
Id & ~RTR - Data Frame Id (~RTR)
Id - Remote or Data Frame Id
Id & RTR - Remote Frame Id (RTR)
All Errors - All Errors
Overload - Overload Frame
Ack Error - Acknowledge Error