Function:span - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:FUNCtion:SPAN

Command Syntax
:FUNCtion:SPAN <span>
<span> ::= the current frequency span in NR3 format. Legal values are
If you set the frequency span to a value outside of the legal range, the
step size is automatically set to the nearest legal value.
The :FUNCtion:SPAN command sets the frequency span of the display (left
graticule to right graticule) when FFT (Fast Fourier Transform) is selected.
Query Syntax
:FUNCtion:SPAN?
The :FUNCtion:SPAN? query returns the current frequency span in Hertz.
After a *RST (Reset) or :AUToscale command, the values returned by the
N O T E
:FUNCtion:CENTer? and :FUNCtion:SPAN? queries depend on the current :TIMebase:RANGe
value. Once you change either the :FUNCtion:CENTer or :FUNCtion:SPAN value, they no
longer track the :TIMebase:RANGe value.
Return Format
<span><NL>
<span> ::= the current frequency span in NR3 format. Legal values are 1
Hz to 100 GHz.
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
1 Hz to 100 GHz.
"Introduction to :FUNCtion Commands"
":FUNCtion:CENTer"
":TIMebase:RANGe"
on page 404
":TIMebase:SCALe"
on page 406
on page 230