5
Commands by Subsystem
:TIMebase:WINDow:SCALe
Command Syntax
:TIMebase:WINDow:SCALe <scale_value>
<scale_value> ::= scale value in seconds in NR3 format
The :TIMebase:WINDow:SCALe command sets the zoomed (delayed)
window horizontal scale (seconds/division). The main sweep scale
determines the range for this command. The maximum value is one half of
the :TIMebase:SCALe value.
Query Syntax
:TIMebase:WINDow:SCALe?
The :TIMebase:WINDow:SCALe? query returns the current zoomed window
scale setting.
Return Format
<scale_value><NL>
<scale_value> ::= current seconds per division for the zoomed window
See Also
•
•
•
•
•
410
(see
page
658)
"Introduction to :TIMebase Commands"
":TIMebase:RANGe"
on page 404
":TIMebase:POSition"
":TIMebase:SCALe"
on page 406
":TIMebase:WINDow:RANGe"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
on page 400
on page 403
on page 409