:FUNCtion:SOURce2

Command Syntax
:FUNCtion:SOURce2 <value>
<value> ::= {CHANnel<n> | NONE}
<n> ::= {{1 | 2} | {3 | 4}} for 4ch models, depending on SOURce1
<n> ::= {1 | 2} for 2ch models
The :FUNCtion:SOURce2 command is only used when an FFT (Fast Fourier
Transform), DIFF, or INT operation is selected (see
the:FUNCtion:OPERation command for more information about selecting an
operation). The :FUNCtion:SOURce2 command selects the source for
function operations. Choose CHANnel<n>, or ADD, SUBT, or MULT to
specify the desired source for function DIFFerentiate, INTegrate, and FFT
operations specified by the :FUNCtion:OPERation command.
If CHANnel1 or CHANnel2 is selected for :FUNCtion:SOURce1, the
SOURce2 selection can be CHANnel1 or CHANnel2. Likewise, if CHANnel3
or CHANnel4 is selected for :FUNCtion:SOURce1, the SOURce2 selection
can be CHANnel3 or CHANnel4.
Query Syntax
:FUNCtion:SOURce2?
The :FUNCtion:SOURce2? query returns the second source for function
operations on two waveforms.
Return Format
<value><NL>
<value> ::= {CHAN<n> | NONE}
<n> ::= {{1 | 2} | {3 | 4}} for 4ch models, depending on SOURce1
<n> ::= {1 | 2} for 2ch models
See Also
242
(see
page
658)
selection
selection
"Introduction to :FUNCtion Commands"
":FUNCtion:OPERation"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
on page 230
on page 237