Overload Event Register (:Ovlregister) - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

Overload Event Register (:OVLRegister)

The Overload Event Register register hosts these bits:
Name
Channel 1 OVL
Channel 2 OVL
Channel 3 OVL
Channel 4 OVL
External Trigger
OVL
Channel 1 Fault
Channel 2 Fault
Channel 3 Fault
Channel 4 Fault
External Trigger
Fault
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
Location
Description
bit 0
Overload has occurred on Channel 1 input.
bit 1
Overload has occurred on Channel 2 input.
bit 2
Overload has occurred on Channel 3 input.
bit 3
Overload has occurred on Channel 4 input.
bit 4
Overload has occurred on External Trigger input.
bit 6
Fault has occurred on Channel 1 input.
bit 7
Fault has occurred on Channel 2 input.
bit 8
Fault has occurred on Channel 3 input.
bit 9
Fault has occurred on Channel 4 input.
bit 10
Fault has occurred on External Trigger input.