Agilent Technologies InfiniiVision 5000 Series Programmer's Manual page 630

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

9
Status Reporting
The next program prints 0xD1 and clears bit 6 (RQS) and bit 4 (MAV) of
the Status Byte Register. The difference in the output value between this
example and the previous one is the value of bit 6 (weight = 64). Bit 6 is
set when the first enabled summary bit is set and is cleared when the
Status Byte Register is read by the serial poll command.
Example
The following example uses the resource session object's ReadSTB method
to read the contents of the oscilloscope's Status Byte Register.
varQueryResult = myScope.IO.ReadSTB
MsgBox "Status Byte Register, Serial Poll: 0x" + Hex(varQueryResult)
Use Serial Polling to Read Status Byte Register. Serial polling is the preferred method to
N O T E
read the contents of the Status Byte Register because it resets bit 6 and allows the next
enabled event that occurs to generate a new SRQ interrupt.
630
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents