Measure:xmax - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

5
Commands by Subsystem

:MEASure:XMAX

Command Syntax
:MEASure:XMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:XMAX command installs a screen measurement and starts
an X- at- Max- Y measurement on the selected window. If the optional
source parameter is specified, the current source is modified.
:MEASure:XMAX is an alias for :MEASure:TMAX.
N O T E
Query Syntax
:MEASure:XMAX? [<source>]
The :MEASure:XMAX? query measures and returns the horizontal axis
value at which the maximum vertical value occurs. If the optional source
is specified, the current source is modified. If all channels are off, the
query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= horizontal value of the maximum in NR3 format
See Also