Operation Status Event Register (:Operegister[:Event]) - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

9
Status Reporting

Operation Status Event Register (:OPERegister[:EVENt])

The Operation Status Event Register register hosts these bits:
Name
RUN bit
WAIT TRIG bit
MTE bit
OVLR bit
HWE bit
If any of these bits are set, the OPER bit (bit 7) of the Status Byte
Register is set. The Operation Status Event Register is read and cleared
with the :OPERegister[:EVENt]? query. The register output is enabled or
disabled using the mask value supplied with the OPEE command.
638
Location
Description
bit 3
Is set whenever the instrument goes from a stop
state to a single or running state.
bit 5
Is set by the Trigger Armed Event Register and
indicates that the trigger is armed.
bit 9
Comes from the Mask Test Event Registers.
bit 11
Is set whenever a 50Ω input overload occurs.
bit 12
Comes from the Hardware Event Registers.
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents