Trigger:uart Commands - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:TRIGger:UART Commands

Table 70 :TRIGger:UART Commands Summary
Command
:TRIGger:UART:BASE
<base> (see
page
:TRIGger:UART:BAUDrat
e <baudrate> (see
page
488)
:TRIGger:UART:BITorde
r <bitorder> (see
page
489)
:TRIGger:UART:BURSt
<value> (see
page
490)
:TRIGger:UART:DATA
<value> (see
page
491)
:TRIGger:UART:IDLE
<time_value> (see
page
492)
:TRIGger:UART:PARity
<parity> (see
page
493)
:TRIGger:UART:POLarit
y <polarity> (see
page
494)
:TRIGger:UART:QUALifi
er <value> (see
page
495)
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
Query
:TRIGger:UART:BASE?
487)
(see
page
487)
:TRIGger:UART:BAUDrat
e? (see
page
:TRIGger:UART:BITorde
r? (see
page
:TRIGger:UART:BURSt?
(see
page
490)
:TRIGger:UART:DATA?
(see
page
491)
:TRIGger:UART:IDLE?
(see
page
492)
:TRIGger:UART:PARity?
(see
page
493)
:TRIGger:UART:POLarit
y? (see
page
:TRIGger:UART:QUALifi
er? (see
page
Options and Query Returns
<base> ::= {ASCii | HEX}
<baudrate> ::= integer from 1200
488)
to 3000000 in 100 b/s increments
<bitorder> ::= {LSBFirst |
489)
MSBFirst}
<value> ::= {OFF | 1 to 4096 in
NR1 format}
<value> ::= 8-bit integer from
0-255 (0x00-0xff) in decimal,
<hexadecimal>, <binary>, or
<quoted_string> format
<hexadecimal> ::= #Hnn where n
::= {0,..,9 | A,..,F} for
hexadecimal
<binary> ::= #Bnn...n where n ::=
{0 | 1} for binary
<quoted_string> ::= any of the
128 valid 7-bit ASCII characters
(or standard abbreviations)
<time_value> ::= time from 1 us
to 10 s in NR3 format
<parity> ::= {EVEN | ODD | NONE}
<polarity> ::= {HIGH | LOW}
494)
<value> ::= {EQUal | NOTequal |
495)
GREaterthan | LESSthan}
Commands by Subsystem
5
485

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents