Mteenable (Mask Test Event Enable Register); Opee (Operation Status Enable Register) - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

4
Commands Quick Reference
Table 3
Root (:) Commands Summary (continued)
Command
:DIGitize
[<source>[,..,<source
>]] (see
page
199)
:MTEenable <n> (see
page
201)
n/a
:OPEE <n> (see
page
205)
n/a
n/a
:OVLenable <mask>
(see
page
211)
n/a
80
Query
n/a
:MTEenable? (see
page
201)
:MTERegister[:EVENt]?
(see
page
203)
:OPEE? (see
page
206)
:OPERregister:CONDiti
on? (see
page
207)
:OPERegister[:EVENt]?
(see
page
209)
:OVLenable? (see
page
212)
:OVLRegister? (see
page
213)
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Options and Query Returns
<source> ::= {CHANnel<n> |
FUNCtion | MATH | SBUS{1 | 2}}
for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | POD{1 | 2} | BUS{1 |
2} | FUNCtion | MATH | SBUS{1 |
2}} for MSO models
<source> can be repeated up to 5
times
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<n> ::= 16-bit integer in NR1
format
<n> ::= 16-bit integer in NR1
format
<n> ::= 15-bit integer in NR1
format
<n> ::= 15-bit integer in NR1
format
<n> ::= 15-bit integer in NR1
format
<mask> ::= 16-bit integer in NR1
format as shown:
Bit Weight Input
--- ------ ----------
10
1024
Ext Trigger Fault
9
512
Channel 4 Fault
8
256
Channel 3 Fault
7
128
Channel 2 Fault
6
64
Channel 1 Fault
4
16
Ext Trigger OVL
3
8
Channel 4 OVL
2
4
Channel 3 OVL
1
2
Channel 2 OVL
0
1
Channel 1 OVL
<value> ::= integer in NR1
format. See OVLenable for <value>

Advertisement

Table of Contents
loading

Table of Contents