Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 209

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

3.
Attach the NW25 tee to the NW25/NW40 adaptor using an NW25 center
ring, and then secure the connection with an NW25 clamp.
4.
Attach a Speedivalve to one of the open ports on the NW25 tee using an
NW25 center ring, and then secure the connection with an NW25 clamp.
5.
Attach an NW25 elbow to the open port on the Speedivalve using an
NW25 center ring, and then secure the connection with an NW25 clamp.
6.
Attach the flanged end of a vacuum hose to the open port on the NW25
elbow, using an NW25 center ring, and then secure the connection with
an NW25 clamp.
7.
Repeat
step 4
through
tee.
8.
Connect the free ends of the vacuum hoses to the roughing pump
vacuum connections on the mass spectrometer's rear panel (see
page
B-8).
Speedivalve
NW25 clamp
NW25/NW40 adaptor
step 6
for the remaining open port on the NW25
Connecting the Edwards oil-free roughing pump
Elbow
Clamps
1-inch ID vacuum hose
B-5