Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 55

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

To install the capillary
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The ESI probe tip is sharp. To avoid puncture wounds, handle
Warning:
the probe with care.
1.
Remove the existing capillary (see
2.
Using the sharp knife or PEEK tubing cutter, cut an approximately
60-cm (24-inches) length of red PEEK tubing.
Requirement:
squarely (that is, perpendicular to the tube's horizontal axis).
3.
Insert one end of the red PEEK tubing in the probe inlet connector, and
finger tighten the connector in the PEEK union.
Doing so ensures a minimum dead volume when fitting the
Rationale:
capillary.
The probe and source components can be contaminated
To minimize dead volume, cut the ends of the tubing
Probe inlet connector
PEEK union
PEEK tubing
Installing the ESI small bore capillary option
page
5-59).
TP02671