3.
Using a long "finger tight" fitting, connect 1/16-inch, 1000-mm,
natural-colored PEEK tubing to port 1 of the sample selector valve, and
thread the tubing into the waste port.
Waste port
Long "finger tight" fitting
The liquid waste system collects waste without requiring a fitted
Tip:
connection. Waste drains through the connector on the left-hand side of
the instrument and into the waste bottle (see
4.
Using a long "finger tight" fitting, connect orange, 1/16-inch, 1000-mm,
PEEK tubing from port 2 of the sample selector valve to the wash bottle.
Push the tubing to the bottom of the wash solution, and
Requirement:
secure it to prevent the tubing floating to the surface during use.