Configuring the LockSpray source
The following table summarizes how you configure the LockSpray source for
the various ionization modes.
Configuring the LockSpray source
Ionization mode
ESI
APCI
ESCi
Configuring for ESI mode
To operate in ESI mode, you must fit the ESI probe to the LockSpray source
enclosure. If you intend using the small-bore capillary option, fit the capillary
to the probe first (see
For more information on using ESI mode, see the SYNAPT G2 MS system
online Help.
Installing the ESI probe
Required materials
•
Chemical-resistant, powder-free gloves
•
PEEK™ tubing
To install the ESI probe
Warning:
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
prepared for working on the source before commencing this procedure.
3-2
Configuring the LockSpray Source
Probe type
ESI
APCI
ESI
page
3-8).
The LC system connections, ESI probe, and source can
To avoid electric shock, ensure that the instrument is
Corona pin fitted?
No
Yes
Yes