To remove the sampling cone assembly from the source
Warning:
Standby mode before commencing this procedure.
Warning:
the source enclosure open if one or both of these conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
Warning:
while working with the source enclosure open.
1.
Close the source isolation valve (see
2.
Grasp the cone gas nozzle handle, and use it to rotate the sampling cone
assembly 90 degrees, moving the handle from the vertical to the
horizontal position.
5-22
Maintenance Procedures
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
To avoid electric shock, ensure that the instrument is in
To avoid puncture wounds, take great care while working with
The source can be hot. To avoid burn injuries, take great care
page
5-15).
Sampling cone
assembly
Cone gas nozzle
handle
TP03131