Deploying The Sprayer Platform Adjuster Assembly - Waters SYNAPT G2 Operator's, Overview And Maintenance Manual

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

NanoLockSpray source configuration
Sprayer type
Universal NanoFlow nebulizer
sprayer
Borosilicate glass capillary
NanoFlow (see
NanoFlow capillary electrophoresis
(CE) sprayer. For details, see the
Capillary Electrophoresis/Capillary
Electrochromatography Sprayer
User's Guide (part number 6666522).

Deploying the sprayer platform adjuster assembly

To move the sprayer platform out of the source
Warning:
over the sprayer.
1.
Confirm that the sprayer's safety cover is installed (see the figure on
page
4-2).
2.
Unscrew the thumbscrew on the front of the sprayer platform.
3.
Pull out the side thumbscrew and withdraw the sprayer platform from
the source.
4.
Release the side thumbscrew, locking the platform in the withdrawn
position.
page
4-8).
To avoid electrical shock, ensure the safety cover is in place
Deploying the sprayer platform adjuster assembly
Used for
Flow injection or for coupling to
nanoACQUITY UPLC with regulated
flow rates down to 100 nL/min.
Extremely low flow rates (less than
100 nL/min).
Stable electrospray by means of a
make-up liquid at the CE capillary
tip. The make-up flow rate is less
than 1 µL/min.
4-5