External connections 2 (continued)
8
Not used
9
Event Out 2 +
10
Event Out 2 -

Signal connections

Mass spectrometer signal connections
Signal connections
Analog Out
Gas Fail Interlock Out Used to stop the solvent flow if the nitrogen gas
Event Out 1, Event
Out 2
Event In 1, Event In 2 Allows an external device to start data acquisition.
CE interlock Out
Requirement:
external electrical disturbances, install connector covers over the signal
connectors.
Required materials
Small flat-bladed screwdriver
Connector covers
B-18
External Connections
Description
Used for analog chart output functionality. The
output voltage range is 0 to 2 V. The resolution of
the voltage output is 12 bits.
supply fails.
Allows the mass spectrometer to trigger an external
event.
Interfaces with a capillary electrophoresis power
supply so that the instrument is safely interlocked
against high voltages.
To meet the regulatory requirements of immunity from
+30 V dc,
100 mA
+30 V dc,
100 mA