Mass spectrometer configuration

The mass spectrometer consists of four principle components: the source, a
quadrupole, a Triwave device, and a TOF mass analyzer. Ionized sample
produced in the source travels through the quadrupole and Triwave. The TOF
detector system records mass spectra as its output.
Using MassLynx and the instrument control software, you control, configure,
and operate the instrument.
The following processes are performed using the MassLynx software:
Configuring the SYNAPT G2 MS system
Calibrating the SYNAPT G2 MS system
Creating inlet and experiment methods that define operating
parameters for an analysis run
Running samples
Monitoring acquisition status
Acquiring data
Processing data
Viewing the data
The MassLynx User's Guide. Also, the MassLynx online Help
See also:
provides more information on installing and using the MassLynx software.

Triwave device

The Triwave device consists of three T-Wave ion guides and is automatically
configured. The device transfers ions from the quadrupole to the TOF with
optimum efficiency, and it provides two separate collision cell regions (TRAP
and TRANSFER T-Waves) for optimized fragmentation of compounds of
interest. The collision energy, and hence the degree of fragmentation, is
manually or automatically controllable.

TOF

With its associated detector, the TOF records mass spectra derived from the
ions' time of flight. A high voltage pulse orthogonally accelerates the ions by
pushing them out across their direction of travel, into a flight tube. A
reflectron reflects the ions back toward the detector.
Mass spectrometer configuration
1-17