To remove the existing capillary
1.
Prepare the instrument for working on the source (see
Warning:
take great care while working with these components.
Warning:
an ESI probe is fitted to the source, remove the probe before
continuing with this procedure.
2.
If an ESI probe is fitted to the source, remove it from the source (see
page
3-7).
3.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
4.
Wait 10 minutes for the source ion block to cool.
5.
Disconnect the "finger tight" nut on the inside of the source enclosure.
LockSpray sprayer
(behind baffle)
The probe and source can be hot. To avoid burn injuries,
The ESI probe tip is sharp. To avoid puncture wounds, if
Replacing the LockSpray probe capillary
page
5-7).
"Finger
tight" nut
5-75