Caution:
sampling cone assembly has been removed from the ion block
assembly.
3.
Slide the sampling cone assembly out of the ion block assembly.
Ion block
assembly
Disassembling the sampling cone assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Do not open the isolation valve at any time when the
The sampling cone assembly can be contaminated with
Cleaning the sampling cone assembly
Combined 2.5-mm
Allen wrench and cone
extraction tool
TP03132
5-23