Replacing the ion block source heater

Replace the ion block source heater if it fails to heat the ion block.
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
Combined 2.5-mm Allen wrench and cone extraction tool
Ion block source heater
To replace the ion block source heater
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Remove the ion block assembly from the instrument (see
2.
Ensure that the isolation valve is closed.
The ion block assembly can be contaminated with
Replacing the ion block source heater
page
5-31).
Isolation valve handle
in closed position
5-87