To use an external reference bottle, push the tubing to the bottom of
the reference solution, and secure the tubing at the bottle neck to
prevent it floating to the surface during use.
5.
Using a long "finger tight" fitting, connect orange, 1/16-inch, 680-mm,
PEEK tubing from port 3 of the LockSpray selector valve, through
tubing guide C and into vial C.
Requirement:
long "finger tight" fitting over it, push the tubing through the left-hand
hole, to the bottom of the vial, and finger tighten the fitting.
6.
Using a long "finger tight" fitting, connect orange, 1/16-inch, 1000-mm,
PEEK tubing from port 6 of the LockSpray selector valve to the wash
bottle.
Requirement:
secure the tubing to prevent it floating to the surface during use.
7.
Using a long "finger tight" fitting, connect 1/16-inch, 1000-mm,
natural-color PEEK tubing to port 5 of the LockSpray selector valve, and
thread the tubing into the waste port.
Waste port
The liquid waste system collects waste without requiring a fitted
Tip:
connection. Waste drains through the connector at the base of the
instrument and into a waste bottle (see
5-104 Maintenance Procedures
As the tubing emerges from the tubing guide, thread the
Push the tubing to the bottom of the wash solution, and
A
A
B
B
C
C
A
Wa ters
C
B
C
B
A
page