Installing the new capillary

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
10-mm wrench
Needle-nose pliers
LC pump
HPLC-grade (or better) 1:1 acetonitrile/water
Capillary
Ferrule
Seal
PTFE liner tubing
Conductive sleeve
Red PEEK tubing
Sharp knife or PEEK tubing cutter
Metal gasket for the probe tip
Safety goggles
To install the new capillary
Warning:
the probe with care.
5-64
Maintenance Procedures
The probe and source components can be contaminated
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The ESI probe tip is sharp. To avoid puncture wounds, handle

Hide quick links: