Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 159

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

To install the new TaperTip and capillary
1.
On the NanoLockSpray reference probe assembly, screw the pin-plug
capillary locator into the end of the union block.
NanoLockSpray
Pin-plug capillary locator
2.
Thread the capillary through the body of the NanoLockSpray reference
probe.
3.
Slide a PEEK coupler over the end of the fused-silica capillary.
4.
Locate the PEEK coupler in the union, and carefully slide the
fused-silica capillary into the union until it butts against the pin-plug
capillary locator.
Doing so ensures a minimum dead volume when fitting the
Rationale:
TaperTip.
5.
Tighten the PEEK coupler to hold the capillary in place.
6.
Unscrew and remove the pin-plug capillary locator.
r
eference probe
Replacing the NanoLockSpray reference probe capillary
5-81