Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 135

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

If the probe tip is difficult to remove, use the 7-mm wrench in
Tip:
conjunction with the 10-mm wrench.
3.
Remove the metal gasket from the probe tip.
Metal gasket
Warning:
contaminated with biohazardous and/or toxic materials.
Dispose of them according to local environmental
regulations.
4.
Dispose of the metal gasket in accordance with local environmental
regulations.
5.
If the probe tip is damaged, dispose of it in accordance with local
environmental regulations.
7-mm wrench
Probe tip
The probe tip and metal gasket can be
Replacing the ESI probe tip and gasket
10-mm wrench
5-57