Replacing the NanoLockSpray reference probe capillary

Replace the NanoLockSpray reference probe TaperTip™ or capillary if it is
irreversibly blocked, or if it is contaminated or damaged.

Removing the NanoLockSpray reference probe

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
5-mm Allen wrench
Source of oil-free, inert gas (nitrogen or argon) for drying (air-drying
optional)
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
Large beaker
To remove the NanoLockSpray reference probe
Warning:
performed on its source before commencing this procedure.
1.
Prepare the instrument for work performed on its source (see
2.
Disconnect the reference probe's cable from the instrument's high
voltage connector.
3.
Disconnect the fused silica capillary from the instrument's IntelliStart
Fluidics system.
4.
Remove the NanoLockSpray source enclosure from the instrument (see
page
5-8).
5.
Retrieve the combined 2.5-mm Allen wrench and cone extraction tool
from its storage location on the source adaptor housing.
5-78
Maintenance Procedures
All plumbing connections, the probe, and the source can
Warning:
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
To avoid electric shock, prepare the instrument for work
page
5-7).