Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 62

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

3.
Tighten the probe locking ring to secure the probe in place.
APCI probe mounted on the source enclosure
Source window
4.
Slide open the instrument's source interface door (see the figure on
page
3-5).
Warning:
to connect the diverter valve to the APCI probe; use the PEEK™
tubing supplied with the instrument.
5.
Using tubing greater than or equal to 0.004-inch ID, connect port 2 (the
top port) of the diverter valve to the APCI probe.
Recommendation:
for sample flow rates ≤1.2 mL/min; use 0.005-inch ID tubing for sample
flow rates >1.2 mL/min.
Requirement:
instrument, minimize the length of the tubing connecting the diverter
valve to the APCI probe. Doing so minimizes delays and dispersion.
3-16
Configuring the LockSpray Source
Vernier probe adjuster
To avoid electric shock, do not use stainless steel tubing
To reduce peak broadening, use 0.004-inch ID tubing
If you are replacing the tubing supplied with the
APCI probe
Probe locking ring
Vertical probe
adjuster
Source enclosure
release
TP03128