1.
Remove the ion block assembly from the PEEK ion block support (see
page
5-31).
2.
Using the 3-mm Allen wrench, unscrew and remove the 4 screws
securing the PEEK ion block support to the source adaptor housing.
Source adaptor
housing
PEEK ion block
support
Securing screws
3.
Remove the PEEK ion block support from the source adaptor housing.
4.
Using the O-ring removal kit, carefully remove all the O-rings from the
PEEK ion block support (see
5.
If any of the O-rings show signs of deterioration or damage, dispose of
them in accordance with local environmental regulations.
6.
Carefully grasp the T-Wave ion guide assembly and remove it from the
source adaptor housing.
5-50
Maintenance Procedures
The O-rings can be contaminated with
Warning:
biohazardous and/or toxic materials. Dispose of them
according to local environmental regulations.
To avoid damage when removing the T-Wave ion guide
Caution:
assembly from the source adaptor housing, avoid scratching the
internal surfaces of the adaptor block.
page
5-17).
T-Wave
ion guide
assembly