a.
Slide the PEEK nut, Super Flangeless ferrule, and stainless steel
ring over the pump end of the steel tubing.
PEEK nut
b.
Insert the tubing in the pump, and tighten the fittings.
c.
At the sample selector valve, screw the natural-color PEEK
female-to-male adaptor into port 7.
d.
Slide the transparent, 1/16-inch flangeless nut and blue, 1/16-inch
flangeless ferrule over the tubing.
Rationale:
case of a blockage.
e.
Insert the tubing in the female-to-male adaptor in port 7, and
tighten the fittings.
5-110 Maintenance Procedures
Stainless steel ring
Flangeless nut
These components provide fail-safe pressure relief in
Super flangeless ferrule
PEEK female -to-male adaptor
Flangeless ferrule
Stainless steel tubing