6.
Rotate the LockSpray sprayer assembly clockwise through 90º, and
remove it from the LockSpray sprayer support assembly.
7.
Using the 4-mm wrench, unscrew and remove the LockSpray probe tip
from the LockSpray sprayer assembly.
8.
Using the 7-mm wrench, unscrew the compression fitting on the rear of
the LockSpray sprayer assembly.
9.
Remove the liner tubing and capillary from the LockSpray sprayer
assembly.
10. Remove the capillary from the liner tubing.
11. Dispose of the capillary in accordance with local environmental
regulations.
5-76
Maintenance Procedures
LockSpray sprayer support assembly
The capillary can be contaminated with
Warning:
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.
LockSpray sprayer assembly