Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 215

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

1.
Locate the nitrogen exhaust trap bottle in an accessible area below the
instrument.
Nitrogen exhaust trap bottle
From instrument
pilot valve port
From instrument
exhaust connection
Nitrogen exhaust
trap bottle
Caution:
tubing squarely (that is, perpendicular to its horizontal axis).
2.
Cut a length of 4-mm tubing long enough to connect the instrument to
the nitrogen exhaust trap bottle.
3.
Connect one end of the tubing to the pilot valve port on the instrument's
rear panel.
4.
Connect the free end of the tubing to the pilot valve port on the nitrogen
exhaust trap bottle (see the figure on
TP03164
To avoid gas leaks, use the sharp knife to cut the PTFE
Connecting the nitrogen exhaust line
To laboratory
exhaust port
Cap
page
B-8).
B-11