Optional Glass Capillary Sprayer; Installing The Glass Capillary Sprayer - Waters SYNAPT G2 Operator's, Overview And Maintenance Manual

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

Optional glass capillary sprayer

The glass capillary sprayer is designed for use with metal-coated borosilicate
glass capillaries, which allow extremely low flow rates (less than 100 nL/min).
The capillaries are used for one sample only and must then be discarded.
To use the glass capillary sprayer, you must complete the following
procedures:
Install the glass capillary sprayer.
Fit and load the glass capillary.
Optimize the sprayer.

Installing the glass capillary sprayer

Required materials
Chemical-resistant, powder-free gloves
To install the glass capillary sprayer
Warning:
biohazardous and/or toxic materials. Always wear
f
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
performed on its source before commencing this procedure.
1.
Prepare the instrument for work performed on its source (see
Warning:
take great care while working with these components.
2.
In the Tune window, click Source Standby
adjacent status display is yellow.
3.
Retract the sprayer platform adjuster assembly from the source (see
page
4-5).
4.
Remove the sprayer's safety cover.
4-8
Configuring the NanoLockSpray source
The source components can be contaminated with
To avoid electric shock, prepare the instrument for work
The probe and source can be hot. To avoid burn injuries,
page
5-7).
, and confirm that the

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents