Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 170

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

1.
Remove the source enclosure from the instrument (see
2.
Using the O-ring removal kit, carefully remove the following seals from
the probe adjuster assembly:
Probe seal
Nebulizer gas seal
3.
Using the O-ring removal kit, carefully remove the following seals from
the source enclosure:
Source enclosure seal
Nebulizer gas seal
Desolvation gas seal
5-92
Maintenance Procedures
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
page
5-8).