1.
Remove the source enclosure from the instrument (see
2.
Using the O-ring removal kit, carefully remove the following seals from
the probe adjuster assembly:
•
Probe seal
•
Nebulizer gas seal
3.
Using the O-ring removal kit, carefully remove the following seals from
the source enclosure:
•
Source enclosure seal
•
Nebulizer gas seal
•
Desolvation gas seal
5-92
Maintenance Procedures
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal
page
5-8).