Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 87

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

4.
Disconnect the source enclosure cables from the instrument's
connectors.
Warning:
take great care while working with the source enclosure open if a
corona pin is fitted.
Caution:
NanoLockSpray source enclosure, move the sprayer platform out
of the source enclosure before opening it (see
5.
Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
6.
Using two hands, grasp the source enclosure, and lift it vertically off the
two supporting studs on the source adaptor housing.
The corona pin tip is sharp. To avoid puncture wounds,
To avoid damaging the sample inlet when removing a
Source enclosure
Removal and refitting of the source enclosure
page
4-5).
Supporting stud
TP03164
5-9