Cleaning The Source T-Wave Ion Guide Assembly; Removing The T-Wave Ion Guide Assembly From The Source Assembly - Waters SYNAPT G2 Operator's, Overview And Maintenance Manual

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

17. Fit the ion block cover plate to the ion block assembly, and then, using
the combined 2.5-mm Allen wrench and cone extraction tool, tighten the
2 captive screws that secure the ion block cover plate.
18. Hold the sampling cone assembly so that the cone gas nozzle handle is
oriented horizontally and at the top, and then slide the sampling cone
assembly into the ion block assembly.
19. Grasp the sampling cone assembly handle, and use it to rotate the
sampling cone assembly through 90 degrees.
20. Fit the ion block assembly to the source assembly (see

Cleaning the source T-Wave ion guide assembly

Clean the source T-Wave ion guide assembly if cleaning the ion block,
isolation valve, and extraction cone fails to increase signal sensitivity.
Removing the T-Wave ion guide assembly from the source
assembly
Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
3-mm Allen wrench
O-ring removal kit
To remove the T-Wave ion guide assembly
The source components can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Cleaning the source T-Wave ion guide assembly
page
5-37).
5-49