Replacing the LockSpray source assembly seals

You do not need to replace the NanoLockSpray source assembly seals;
Note:
this section relates to the LockSpray source only.
Mechanical wear and solvent degradation of the source assembly seals can
cause gas leakage from the source. Annual replacement of the following seals
reduces the chance of these seals failing:
Probe adjuster assembly probe seal
Probe adjuster assembly nebulization gas seal
Source enclosure seal
Source enclosure nebulizer gas seal
Source enclosure desolvation gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
Chemical-resistant, powder-free gloves
O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The source components can be contaminated with
Replacing the LockSpray source assembly seals
5-91