Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 157

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

6.
Using the combined 2.5-mm Allen wrench and cone extraction tool,
unscrew the 3 fixing screws from the top of the NanoLockSpray
reference probe.
Warning:
of exposed, fused-silica TaperTip, which is sharp and fragile. To
avoid puncture wounds, handle the probe with care.
7.
Remove the NanoLockSpray reference probe from the probe adjuster
assembly.
8.
Unscrew the TaperTip PEEK coupler, and remove the TaperTip from
the union.
PEEK couplers
TaperTip
The NanoLockSpray reference probe tip consists of 5 cm
Fused-silica capillary
Union
Replacing the NanoLockSpray reference probe capillary
Fixing screw
NanoLockSpray reference
probe
Fused-silica capillary with
PEEK protective sleeve
5-79