Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 151

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

9.
Perform a leak test by attaching the free end of the PEEK tubing to an
LC pump and pumping 50:50 acetonitrile/water through it, at 1 mL/min.
If leakage occurs, disassemble and remake the connection, and
repeat the leak test.
If the backpressure on the LC pump is high, replace the capillary,
and repeat the leak test.
10. When no leakage occurs and the backpressure on the LC pump is
normal, disconnect the PEEK tubing from the LC pump.
11. Remove the probe inlet connector and PEEK tubing from the PEEK
union.
12. Remove the probe heater (see
13. Fit the PEEK union/UNF coupling assembly to the nebulizer adjuster
knob.
14. Carefully thread the capillary through the probe assembly.
15. Carefully push the PEEK union/UNF coupling assembly and capillary
into the probe assembly so that the locating pin on the UNF coupling is
fully engaged in the locating slot at the head of the probe assembly.
UNF coupling locating pin
16. Fit the nebulizer adjuster knob to the PEEK union/UNF coupling
assembly.
17. Finger-tighten the nebulizer adjuster knob onto the probe assembly.
18. Fit the probe gasket and end cover to the probe assembly.
19. Using the combined 2.5-mm Allen wrench and cone extraction tool, fit
and tighten the 3 screws that secure the end cover.
page
5-84).
Probe assembly locating slot
Replacing the APCI probe sample capillary
5-73