Removing the ESI probe tip and gasket

Required materials
Chemical-resistant, powder-free gloves
7-mm wrench
10-mm wrench
To remove the ESI probe tip and gasket
Warning:
great care while performing this procedure.
Warning:
the probe with care.
1.
Remove the ESI probe from the source (see
2.
Using the 10-mm wrench, remove the probe tip.
5-56
Maintenance Procedures
The probe and source components can be contaminated
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The probe and source can be hot. To avoid burn injuries, take
The ESI probe tip is sharp. To avoid puncture wounds, handle
page
3-7).
10-mm wrench
Probe tip