Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 179

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

Required materials
Chemical-resistant, powder-free gloves
SYNAPT G2 fluidics tubing and fitting kit
This kit contains components for both the sample and LockSpray
Tip:
system plumbing.
For the LockSpray reference probe connection
375 mm of 0.005-inch ID red PEEK
Long "finger tight" fitting
For the NanoLockSpray source reference probe connection
375 mm of 25-µL, fused-silica capillary (supplied in 1-m lengths)
Protective sleeve consisting of 0.015-inch ID PEEK tubing
F-130 "finger tight" nut
To plumb the LockSpray system
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while working on the
system.
1.
Open the access doors to the IntelliStart Fluidics system.
2.
Using a PEEK nut, Super Flangeless™ ferrule, and stainless steel (SS)
ring, connect the blue, 1/16-inch, 300-mm, PEEK tubing from the
LockSpray pump to port 7 on the LockSpray selector valve, where you
use the long "finger tight" fitting.
All tubing and fittings can be contaminated with
Replacing the IntelliStart Fluidics tubing 5-101