Calibration Information; Flow Rates For The Acquity Uplc Synapt G2 Ms Uplc/Ms/Ms System - Waters SYNAPT G2 Operator's, Overview And Maintenance Manual

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

5.
Start the MassLynx software.
You can monitor the Instrument Console for messages and LED
Tip:
indications.
6.
Click IntelliStart, in the MassLynx main window's lower, left-hand
corner.
Result:
spectrometer is in Standby mode.
7.
Click Operate
Result:
software displays "Ready" in the Instrument Console.

Calibration information

You must calibrate the mass spectrometer prior to use. You can perform this
task using IntelliStart software.
The mass spectrometer's online Help.
See also:
Flow rates for the ACQUITY UPLC SYNAPT G2 MS UPLC/MS/MS
system
The ACQUITY UPLC system can run at high flow rates. To optimize
desolvation, and thus sensitivity, run the ACQUITY UPLC SYNAPT G2 MS
UPLC/MS/MS system at appropriate gas flows and desolvation temperatures.
Flow rate versus temperature and gas flow
Flow rate
(mL/min)
0.000 to 0.020
0.020 to 0.100
0.101 to 0.300
0.301 to 0.500
>0.500
The mass spectrometer's console appears. The mass
.
When the mass spectrometer is ready to operate, IntelliStart
Source
temperature (°C)
100
120
120
150
150
Desolvation
temperature (°C)
200
350
450
500
600
Starting the mass spectrometer
Desolvation gas
flow (L/h)
800
800
800
1000
1200
2-3