Removing the IntelliStart Fluidics tubing

This procedure explains how to remove the IntelliStart Fluidics tubing and
disconnect the probe tubing at the diverter valve or grounded union.
Required materials
Chemical-resistant, powder-free gloves
To remove the tubing
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while working on the
system.
1.
Open the access doors to the IntelliStart Fluidics system.
2.
Unscrew and remove the PEEK thumbscrews and tubes from the
LockSpray selector, sample selector and diverter valves.
3.
Unscrew and remove the PEEK thumbscrews and tubes from between
the flow sensor and the grounded union.
4.
Close the access doors to the IntelliStart Fluidics system.
5.
Dispose of the tubes and fittings in accordance with local environmental
regulations.

Plumbing the IntelliStart Fluidics LockSpray system

This section describes how to plumb the LockSpray system.
Requirement:
All tubing and fittings can be contaminated with
The IntelliStart Fluidics tubing and fittings can be
Warning:
contaminated with biohazardous and/or toxic materials.
Dispose of them according to local environmental
regulations.
Finger tighten all PEEK fittings.
Replacing the IntelliStart Fluidics tubing
5-99