Removing the ESI probe

Required materials
Chemical-resistant, powder-free gloves
To remove the ESI probe
Warning:
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
performed on its source before commencing this procedure.
1.
Prepare the instrument for work performed on its source (see
2.
Disconnect the fluidics tubing from the ESI probe.
3.
Disconnect the ESI probe's cable from the high voltage connector.
4.
Unscrew the probe locking ring.
Warning:
handle the probe with care.
5.
Carefully remove the ESI probe from the probe adjuster assembly.
6.
If available, fit the protective sleeve to the ESI probe tip.
The LC system connections, ESI probe, and source can
To avoid electric shock, prepare the instrument for work
The ESI probe tip is sharp. To avoid puncture wounds,
page
Configuring for ESI mode
5-7).
3-7