19. Fit the new metal gasket to the probe tip.
Metal gasket
20. Fit the probe tip over the capillary, and screw the tip onto the probe
assembly.
Caution:
21. Using the 10-mm wrench, tighten the probe tip.
22. Using the nebulizer adjuster knob, adjust the capillary so that it
protrudes by approximately 0.5 mm from the end of the probe tip.
During normal operation, the adjuster knob relies on gas pressure
Tip:
to retract the capillary. To retract the capillary in the absence of gas
pressure, invert the probe, and use gravity.
3-12
Configuring the LockSpray Source
To avoid gas leakage, fully tighten the probe tip.
10-mm wrench
Probe tip