Waters SYNAPT G2 Operator's, Overview And Maintenance Manual page 190

Hide thumbs Also See for SYNAPT G2:
Table of Contents

Advertisement

5.
Using a long "finger tight" fitting, connect orange, 1/16-inch, 680-mm,
PEEK tubing from port 4 of the sample selector valve, through tubing
guide A and into vial A.
Tubing guide A
A
B
C
Vial A
Requirement:
long "finger tight" fitting over it, push the tubing through the right-hand
hole, to the bottom of the vial and tighten the fitting.
Follow these suggestions if you find threading the tubes through
Tips:
the guides difficult:
Thread the tubes upward from the bottom hole.
Using needle-nose pliers, make a 20° bend, 10 mm from the end of
the tubing. Then rotate the tubing, as necessary, as you thread it
through the guide.
6.
Repeat this procedure for port 5 to vial B and port 6 to vial C.
5-112 Maintenance Procedures