Operation Of The Keypad; Install Freezer; Choose Location - Thermo Scientific Revco 8920 Operating Manual

-86c ult upright freezer
Table of Contents

Advertisement

Operation of the

Install Freezer

Choose Location

Thermo Scientific
Revco ExF freezer has five basic modes which allow freezer setup and
operation. Press the Mode key to scroll through the mode selections.
Keypad
Note If tipped more than 45°, allow the unit to sit upright for 24 hours
before start up.
To remove the freezer from the pallet, use a 7/16" wrench to remove all
the bolts securing the shipping bracket to the pallet.
Remove the shipping bracket. Remove the ramp boards from the pallet
and place the slotted end over the ramp brackets on the pallet. The
support blocks on the ramps will be facing down. Before moving the
freezer, make sure the casters are unlocked and moving freely. Align the
caster with the ramp boards. Use adequate personnel to roll the freezer off
the pallet.
The freezer can be easily pushed to the desired approved location,
described below. If necessary, the doors and lower front panel may be
opened to move the unit through tight openings. When the freezer is in
position, set the front caster brakes.
Note The freezer must not be moved with the product load inside.
Note For proper ventilation and airflow, a minimum clearance of 5" at the
rear and top and a clearance of 8" on the side of the freezer is required.
Allow adequate space in the front of the freezer for door opening.
Locate the freezer on a firm, level surface in an area with an ambient
temperature between 18°C and 32°C. Provide ample room to reach the
mains disconnect switch (power switch) located on the rear of the freezer.
Up Arrow: Increases or toggles the parameter value.
Enter: Must press Enter key to save to memory all
changed values.
Down Arrow: Decreases or toggles the parameter value.
Silence Key: Press to silence the audible alarm. See
Section 4 for alarm ringback times.
Scroll for Parameters Arrows: Steps the operator through
the parameters of SETTINGS, CALIBRATE and
CONFIGURATION Modes. Right arrow goes to the next
parameter, left arrow returns to previous parameter.
s
Section 1
Installation and Start-Up
s
s
Revco ExF