Table 5.3. G3Lite 5-pin Eurofast connector description.
Pin 1**
Pin 2**
Pin 3
Pin 4
Pin 5
*If using a 6-wire load cell, connect "Sense +" to "Excitation +" and "Sense –" to "Excitation –"
**If using the optional Alarm Relay, pins 1 and 2 are dry contact
Thermo Scientific

5.2 Connector pin configurations

The G3Lite has both 5-pin and 8-pin Eurofast
numbering is shown for the mounted female receptacles.
3
2
Figure 5.2. 5-pin connector.
AUX
AI
Setpoint +
+24V
Setpoint return
24V return
Process value +
Process value +
Process value
Process value
return
return
Shield
Shield
4
5
1
Temp
DO
RTD return
DO cathode
RTD S
Therm–
Drain wire coupled
N/C
to DO signal
RTD F
DO anode
Drain coupled
Therm+
to braid
HyPerforma G3Lite Bioreactor Controller User's Guide
connectors. Pin
Load cell*
Comm.
Signal –
Shield
5V Excitation +
N/C
Signal +
N/C
Tx/Rx –
5V Excitation –
(blue)
TR/Rx –
Shield
(white)
|